Transesophageal echocardiographic diagnosis of a WATCHMAN left atrial appendage closure device thrombus 10 years following implantation.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Additional Information
    • Abstract:
      Left atrial appendage closure with the WATCHMAN device is an alternative to chronic oral anticoagulation for thromboembolic prophylaxis in atrial fibrillation patients. Left atrial device-related thrombus ( DRT) has been described in the first year after implant with an incidence of ~6%. A 79-year-old man underwent WATCHMAN device placement in 2006. Routine protocol specified follow-up transesophageal echocardiograms ( TEE) at 6 weeks, 6 months, and 1 year following implant showed no evidence for DRT or peri-device flow. A decade after device implant, the patient presented with severe symptomatic aortic stenosis and underwent repeat TEE, which revealed a 21 mm × 18 mm DRT on the LA aspect of the WATCHMAN device. He was prescribed apixaban 5 mg po BID. A TEE performed 111 days later demonstrated marked diminution in the DRT (9 mm in diameter). This case demonstrates that WATCHMAN DRT may occur late following implantation. [ABSTRACT FROM AUTHOR]
    • Abstract:
      Copyright of Echocardiography is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)