    • Abstract:
      Background: Four oscillometric devices for self-measurement of blood pressure (SBPM) were evaluated according to the European Society of Hypertension (ESH) international protocol and its 2010 revision in four separate studies. The Omron® M2, Omron M3, and Omron M6 measure blood pressure (BP) at the brachial level, while the Omron R2 measures BP at the wrist level.
      Methods: The international protocol requires a total number of 33 subjects in which the validation is performed. The Omron M2 and Omron R2 were validated in 2009 according to the ESH international protocol, while the Omron M3 and Omron M6 were validated in 2010-2011 according to the 2010 ESH international protocol revision. The protocol procedures were followed precisely.
      Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were 2.7 ± 5.0 and -1.4 ± 3.2 mmHg for systolic and diastolic BP, respectively, using the Omron M2 device, and 1.7 ± 3.2 and -0.9 ± 2.6 mmHg using the Omron M3, 1.6 ± 2.9 and -0.9 ± 2.5 mmHg using the Omron M6, and -1.1 ± 4.8 and -0.9 ± 4.3 mmHg using the Omron R2.
      Conclusion: Readings from the Omron M2, Omron M3, Omron M6, and Omron R2, differing by less than 5, 10, and 15 mmHg, fulfill the ESH international protocol and its 2010 revision requirements. Therefore, each of these four devices can be used by patients for SBPM.
    • References:
      Blood Press Monit. 2010 Feb;15(1):23-38. (PMID: 20110786)
      Blood Press Monit. 2002 Feb;7(1):3-17. (PMID: 12040236)
      J Hypertens. 2007 Jun;25(6):1105-87. (PMID: 17563527)
      J Hypertens. 2005 Apr;23(4):697-701. (PMID: 15775768)
      J Hypertens. 1993 Jun;11(6):677-9. (PMID: 8397248)
      J Hypertens. 2008 Aug;26(8):1505-26. (PMID: 18622223)
      Hypertension. 2005 Jan;45(1):142-61. (PMID: 15611362)
      J Hypertens. 2000 May;18(5):493-508. (PMID: 10826551)
      Hypertension. 2003 Dec;42(6):1206-52. (PMID: 14656957)
    • Contributed Indexing:
      Keywords: European Society of Hypertension; M2; M3; M6; Omron R2; blood pressure measurement; international protocol; validation
    • Publication Date:
      Date Created: 20111217 Date Completed: 20120709 Latest Revision: 20240314
    • Publication Date:
      20240314
    • Accession Number:
      PMC3237100
    • Accession Number:
      10.2147/VHRM.S27193
    • Accession Number:
      22174581