      The varied line spacing plane grating monochromator (VLS-PGM) has been constructed for an industry application beamline using soft x-rays from the 1.4 GeV storage ring in Saga, which is the first synchrotron in Kyushu Island. The design aim of the VLS-PGM is to cover the wide-energy range from 40–1,000 eV and also to have medium resolving power, high throughput, better reproducibility and suppression of higher-order light. In order to cover the wide energy range with suppressing higher-order light, a set of three focusing mirrors and one grating is installed. The resolving power expected from the ray-tracing calculation using SHADOW is more than 5,000 with 10 micron slits. The conventional photoelectron spectrometer is installed with an ion gun, a flood gun and a twin-anode x-ray source. The instrument for XAFS measurement is also attached to the main chamber. © 2007 American Institute of Physics [ABSTRACT FROM AUTHOR]
    • Abstract:
      Copyright of AIP Conference Proceedings is the property of American Institute of Physics and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)